Product No
FY-ER3602Sample Type
Serum, Plasma, Tissue homogenate and Other biological samplesAssay Time
3.5H
              
              Contact supplier to negotiate price details
            
              
              100% Quality Assurance
            
              
              Fast Shipping
            
          
        
                                      | Concentration | OD1 | OD2 | Average OD | Corrected | 
|---|---|---|---|---|
| 500 | 2.498 | 2.508 | 2.503 | 2.435 | 
| 250 | 1.642 | 1.644 | 1.643 | 1.579 | 
| 125 | 0.976 | 0.984 | 0.98 | 0.913 | 
| 62.5 | 0.457 | 0.458 | 0.458 | 0.394 | 
| 31.25 | 0.248 | 0.25 | 0.249 | 0.185 | 
| 15.63 | 0.15 | 0.15 | 0.15 | 0.087 | 
| 7.82 | 0.098 | 0.099 | 0.099 | 0.035 | 
| 0 | 0.063 | 0.063 | 0.063 | 0 | 
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 95-103 | 88-98 | 95-103 | 
| Average (%) | 97 | 98 | 101 | |
| 1:4 | Range (%) | 87-98 | 87-103 | 80-97 | 
| Average (%) | 88 | 91 | 91 | |
| 1:8 | Range (%) | 85-92 | 89-97 | 93-102 | 
| Average (%) | 87 | 96 | 94 | |
| 1:16 | Range (%) | 80-96 | 85-94 | 87-103 | 
| Average (%) | 82 | 89 | 98 | 
| Item | Intra-assay Precision | Intra-assay Precision | 
|---|---|---|
| Sample | 3 | 3 | 
| Replicate | 9 | 18 | 
| CV(%) | 5 | 8 | 
| Sample Type | Range(%) | Average Recovery(%) | 
|---|---|---|
| Serum (n=8) | 87-95 | 91 | 
| EDTA plasma (n=8) | 96-107 | 102 | 
| Cell culture media (n=8) | 80-96 | 87 |